    【Case Study】Infrared Thermal Imaging System Registration in China

    from CIRS by

    1. Introduction

    Product Name: Infrared Thermal Imaging System, Medical Infrared Thermograph, Thermal Texture Maps System, Infrared Fluorescence Positioning Observation Camera

    Product Composition: Generally, it consists of an infrared camera, a processing system, a stand and a display screen. The thermal image of the human body is marked by infrared photography.

    Working Principle: The infrared radiation of the human body is collected by a thermal imaging system and converted into a digital signal to form a pseudo color heat map.

    Mechanism of ActionNot Applicable

    2. Classification

    Classification Code: 06-13-01 Class II Thermal Infrared Imager

    3. Registration Unit

    1) In principle, the attachment of active medical devices and the host used for connection shall be declared as the same registration unit. For the separately registered medical device management accessories, the accessories with different expected USES are divided into different registration units in principle, and the active and passive accessories are divided into different registration units in principle. If the active and passive attachments are in the same aseptic package, they are in principle divided into the same registration unit.

    2) Different types of medical devices with the same scope of application, product performance and structural composition are divided into the same registration unit in principle. However, if the models differ greatly in terms of application, performance, and structure, they are classified into different registration units.

    4. Technical Requirements

    1. Basic technical performance

    1.1 Working band: 8μm~14μm, tolerance ±20%.--Infrared band emitted by human body

    1.2 Instantaneous field of view: better than 1.1mrad.--The instantaneous field of view is the light receiving angle or observation field of a single detector in the sensor. This area is the smallest unit that the sensor can resolve. The smaller the IFOV, the smaller the minimum resolvable unit and the higher the spatial resolution of the image. The IFOV depends on the size of the sensor optics and detector.

    1.3 Field of view: 20o(HX15o(V, tolerance ±20%; - The field angle determines the field of view of the optical instrument. The larger the field angle, the larger the field of view and the smaller the optical magnification. In other words, the target object will not be captured in the lens beyond this angle.

    1.4 Working distance: 0.6m~5m.

    1.5 Image field period: 60Hz, tolerance ± 2Hz .--Infrared camera camera sampling frame frequency, the number of samples extracted from continuous signals and composed of discrete signals per second

    1.6 Warm-up time: The image generation time is not more than 60s, and the time for ensuring the accuracy of temperature measurement is not more than 30min.

    1.7 Focusing function:

    The system has a manual focus adjustment function, which can adjust the image to the clearest target within working distance.

    2. Temperature measurement performance

    2.1 Temperature measurement range: 10 ° C ~ 42 ° C;

    2.2 Temperature measurement accuracy: δ ≤ 0.4 ° C;

    2.3 NETD: 0.1 ° C;

    2.4 temperature measurement repeatability: σ ≤ 0.2 ° C;

    Table 1 Relevant Product Standards

    Standard Number

    Standard Name

    GB 9706.12007

    Medical electrical equipment - Part 1: General requirements for safety

    GB/T 1912008

    Packaging, storage and transportation icon

    GB/T 147102009

    Medical electrical environment requirements and test methods

    YY/T 03162016

    Medical device risk management for medical devices

    YY 05052012

    Medical electrical equipment - Part 1-2: General standard for safety. Parallel standard: Electromagnetic compatibility requirements and testing

    YY/T 0466.12009

    Medical devices - Symbols for labeling, marking and providing information on medical devices - Part 1: General requirements

    YY(/T): Industry standard (Recommendation)

    GB(/T): National standard (Recommendation)

    5. Clinical Trials

    Clinical Trials: not required for the following

    In ‘Medical Device Catalog Exempted from Clinical Trials’

    06-13-01 Thermal Infrared Imager

    Generally, it consists of an infrared camera, a processing system, a stand and a display screen. The thermal image of the human body is marked by infrared photography.

    6. Registration Cost and Duration

    Product Testing :

    Test items

    Testing duration

    Testing fee(USD)

    Safety Performance

    60 working days

    5,700

    EMC

    3,600

    Notes:

    1) Technology Approval

    60 working days

    Another 60 working days if the 1-year supplement is required

    2) Administrative Approval

    20 working days

      