• <blockquote id="sqo4s"></blockquote>
  • <rt id="sqo4s"><kbd id="sqo4s"></kbd></rt>
  • <td id="sqo4s"></td>
    Chemicals
    CIRS Group
    Chemicals
    Cosmetic
    Food
    Medical Devices
    Agrochemicals
    C&K Testing
    Carbon Neutrality
    Search

    Test requirements of existing disinfection product (Biocides) in China

    from CIRS by

    Background

    In accordance with Regulations on the Safety Evaluation Report of Disinfecting Products (2014), disinfection product should perform required tests to complete Safety Evaluation Report. Those testing should be conducted in local Disease Prevention and Control Centre (provincially). There are five Categories of disinfection products and each of them has specific test requirements.

    Category I: Disinfectant


    Test Item

    Disinfection objects

    Skin

    Mucosa

    Hand

    Tableware

    Fruits and vegetables

    Drinking water

    Swimming pool water

    Hospital waste water

    Air

    Medical device and consumable goods

    General Object surface and textile fabrics

    Others

    Sterilization and high level disinfection

    Mediate level disinfection

    Low level disinfection

    Appearance

    Active ingredient content

    pH value①

    Stability

    Stability in continuous use

    ±

    ±

    ±

    ±

    Pb、As、Hg content②

    ±

    Mental corrosion test

    ±

    ±

    ±

    ±

    ±

    Laboratory determination of microbial killing effect③④⑤

    Simulated field tests or field trials⑥

    Toxicological safety testing⑦

    ±

    Overall performance test

    Note: “+” means mandatory item, “-” means exempted item, “±” means optional item.

    ①Glutaraldehyde disinfectant shall measure the pH values before and after adding pH conditioning agent. If the product is solid, perform the test at the highest concentration.

    ②Tableware, drinking water, and fruits and vegetables only conduct Pb and As content test.

    ③Microbial killing test should be based on application scope and valid microbe written on the label and specification.

    ④For Ethanol disinfectant, glutaraldehyde disinfectant, sodium hypochloride disinfectant and bleaching powder disinfectant, if they are used for general object surfaces and textile fabrics, select staphylococcus aureus to conduct quantitative killing test; if they are used for sanitary wares, select candida albicans to conduct quantitative killing test; If they are used at drinking water, swimming pool water or waste water, should choose coli to perform quantitative test; if they are used for tableware, please conduct poliovirus quantitative inactivating test; if they are used for hands, skins or mucosa disinfection, please conduct candida albicans quantitative killing test; if they are used for medical devices and consumable goods sterilization or high level disinfection, please conduct bacterial spores quantitative killing test; if they are used for hands, skins or mucosa disinfection, please conduct candida albicans quantitative killing test while mediate level disinfection shall perform mycobacterium chelonae quantitative killing test; if they are used for air disinfection, please conduct staphylococcus albus quantitative killing test; other type disinfectants should choose a most resistant microorganism, which is decided by product label and specification, to perform efficacy test.

    ⑤The interfering substance concentration is 0.3% for sodium hypochloride disinfectant and disinfectant used after cleaning.

    ⑥The microbe used for medical device and consumable goods disinfectant (includes the disinfectant carried by nonwovens) simulated field test should choose the most resistant microbe according to application scope.

    ⑦Except Ethanol disinfectant, glutaraldehyde disinfectant, sodium hypochloride disinfectant and bleaching powder disinfectant, other disinfectants are required to conduct an acute oral toxicity test or an acute inhalation toxicity test and a mutagenicity test; if the disinfectant is used for hand or skin disinfection, more than once skin irritation trials are required; if the disinfectant is used at mucosa or broken skin, an eye irritation trail is required, if the disinfectant is used for vagina disinfection, a vagina irritation test is required.

    Category II: Disinfection Device/Equipment


    Test Items

    Disinfection objects

    Skin

    Mucosa

    Hand

    Tableware

    Fruits and vegetables

    Drinking water

    Swimming pool water

    Hospital waste water

    Air

    Medical device and consuming goods

    General Object surface and textile fabrics

    Others

    Sterilization and High level disinfection

    Mediate level disinfection

    Low level disinfection

    Efficacy test of main disinfecting factor (includes changing curve)①

    ±

    Pb、As、Hg content test(Only for chemical disinfecting factor device) ②

    ±

    Metal corrosion test(Only for chemical disinfecting factor device)) ③

    ±

    ±

    ±

    ±

    ±

    ±

    Microorganism efficacy test in the lab④

    Simulated field tests or field trials⑤

    Toxicological safety testing⑥

    Overall performance test

    Note: “+” means mandatory item, “-” means exempted item, “±” means optional item.

    ①Except ethylene oxide disinfection (sterilization) cabinet, low temperature plasma sterilization device and LTSF sterilization cabinet, other disinfection device must conduct this test.

    ②Tableware, fruits and vegetables, and drinking water are only required to perform As, Pb content test.

    ③If the chemical factor disinfection device doesn‘t indicate it cannot be used at metal object in label or commercial specification, it must complete metal corrosion test.

    ④Ultraviolet lamp are not required to do efficacy test while other disinfectants conduct efficacy test .One-star tableware disinfection cabinet should conduct a coli quantitative killing test; two-star tableware cabinet should conduct a poliovirus quantitative killing test; pressure steam sterilizer should conduct a Bacillus stearothermophilus spore quantitative killing test.

    ⑤Simulated field test uses the most resistant microbe in the application scope.

    ⑥The disinfection device which generates disinfectant (except sodium hypochloride) should conduct an acute toxicity test or an acute inhalation toxicity test and a mutagenicity test; if the disinfection device is used for hand or skin disinfection, more than once skin irritation trials are required; if the disinfectant is used at mucosa or broken skin, an eye irritation trail is required.

     

    Category III: Chemical Indicator&Biological Indicator

    Test Items

    Indicator card for Ultraviolet lamp radiation intensity

    Disinfectant concentration test paper

    Biological indicator

    Sterilization chemical indicator③

    Microbe contents of BI

    Survival time and killing time

    D value

    Color change of CI in the corresponding sterilization situation①

    Influence factor

    Microorganism survival amount in corresponding disinfecting factor situation②

    Ultraviolet intensity test

    Disinfectant concentration comparison

    Stability

    Others

    ±

    ±

    Note: “+” means mandatory item, “-” means not required item, “±” means optional item.

    ①Includes a successful test and a failed test。

    ②Moist beat disinfection, low temperature hydrogen peroxide gas plasma sterilization, low temperature steam formaldehyde(LTSF) sterilization should use bacillus stearothermophilus spores as tested bacteria. Other disinfection factors should choose Bacillus subtilis variant as tested bacteria.

    ③Includes chemical indicators for sterilizing effect and chemical indicator for sterilizing process.

     

    Table IV: Packaging materials for disinfected and sterilized product

    Test Items

    Packaging material

    Paper

    Non-paper

    Breathable material

    Non-breathable material

    General check for packaging material

     

    Packaging material sterilization period of validity test

     

    Packaging material quality test

     

    Sterilizing factor penetration test

     

    Influence test for sterilizing on packaging label

     

    Packaging material gas impermeability test

     

    Microbial barrier test for breathable mateial

     

    Microbial permeability test

     

    Shelf life test for packaging material

    Note: “+” means mandatory item, “-” means exempted item, “±” means optional item.

     

    Category V: Antibacterial and Bacteriostatic Agent

    Test Items

    Antibacterial Agent

    Bacteriostatic Agent

    Active ingredient content①

    Stability

    pH Value②

    Microbial index:

    Total bacterial colonies number

    Coliform

    Total fungal colonies number

    Pathogenic purulent bacteria

     

     

    Killing microbial index

    Killing test for coli

    Killing test for staphylococcus aureus

    Killing test for candida albicans③

    Killing test for other microbe④

    Bacteriostatic index:

    Bacteriostatic test for coli

    Bacteriostatic test staphylococcus aureus

    Bacteriostatic test candida albicans③

    Bacteriostatic test other microbe④

     

    ±

    ±

     

     

     

    ±

    ±

    Toxicity indicator test⑤

    Note: “+” means mandatory item, “-” means not required item, “±” means optional item.

    ①Only for chemical ingredient.

    ②Except cream products

    ③ The product which label or specification claims it is valid for fungi or used at genitals shall test this item.

    ④If label or specification claims the product can kill or inhibit a certain microbe, the product shall perform this test.

    ⑤If the label or specification claims the antibacterial or bacteriostatic agent used for skin, it should conduct skin irritation trail repeatedly; if the product requires rinsing after use, it just need to conduct acute skin irritation test which exposure time is 2 hours; If the label or specification claims the antibacterial or bacteriostatic agent used for mucosa, it should conduct eye irritation trail; If the label or specification claims the antibacterial or bacteriostatic agent used for vagina mucosa, it should conduct vagina mucosa irritation test.

    Contact Us

    If you need further information and/or still have some questions, please contact us at service@cirs-group.com for more information.

     

     

      

    We have launched a LinkedIn newsletter to keep you up to date on the latest developments across the chemical industry including food and FCMs and personal and home care.

  • <blockquote id="sqo4s"></blockquote>
  • <rt id="sqo4s"><kbd id="sqo4s"></kbd></rt>
  • <td id="sqo4s"></td>
    丝瓜视频